For the detection of samples that are not UV absorption!
Available in 3 different configurations!
Baseline stabilization is improved compared with existing products!
High reproducibility is observed on repeated measurements!
Sample : Polystyrene
Rapid and high performance analysis!!
Solvent usage drastically reduced!!
Usable in ultra rapid SEC columns!!
High reproducibility!!
Type | Analytical | Semi-Preparative | Semi-micro |
---|---|---|---|
Model | RI-501 | RI-502 | RI-504 |
Measuremennt method | Deflection type | ||
Flow cell | 2chamber type | ||
Refractive Index range | 1.00-1.75 | ||
Linearity range | 600 μRIU | 6000 μRIU | 600 μRIU |
Recorder output range selection | (0.25-512) μRIU | (2.5-5120) μRIU | (0.25-512) μRIU |
Integrator output range selection | 128/512 μRIU | 1280/5120 μRIU | 128/512 μRIU |
Noise | 2.5 nRIU (Water_0 mL/min, response;1.5 Sec) |
25 nRIU (Water_0 mL/min, response;1.5 Sec) |
5 nRIU (Water_0 mL/min, response;1.5 Sec) |
Drift | 0.2 μRIU/H | 2 μRIU/H | 0.2 μRIU/H |
Response | 0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec | ||
Auto zero | Full auto zero | ||
Auto zero range | All Refractive Index Range | ||
Integrator output | (0-1) V | ||
Recorder output | (0-10) mV | ||
External Signal Input | Zero/Purge | ||
External Signal Output | READY/LEAK/ERROR | ||
Flow cell volume | 8 μL | 8 μL | 2.5 μL |
Temperature! control system | OFF, (30-55) ℃ (Double Temperature control) | ||
Usable flowrate | (0.2-3) mL/min | (1-50) mL/min | (0.2-1) mL/min |
Maximum back pressure | 50 kPa | ||
Dead volume | IN→Cell; 60 μL | IN→Cell; 120 μL | IN→Cell; 10 μL |
Cell→OUT; 630 μL | Cell→OUT; 540 μL | Cell→OUT; 385 μL | |
Communication port | LAN | ||
Power source | AC(100-240) V±10 %,50/60 Hz 150 VA | ||
Dimensions and Weight | 260 W X 420 D X 165 H(mm) 12.5 kg |